Analysis of temperature rise on the surface of buchanan plugger using thermocouple
Jin-Suk Cho, Yun-Chan Hwang, Sun-Ho Kim, In-Nam Hwang, Bo-Young Choi, Young-Jin Jeong, Woo-Nam Juhng, Won-Mann Oh
J Korean Acad Conserv Dent 2003;28(4):334-340.   Published online July 31, 2003
DOI: https://doi.org/10.5395/JKACD.2003.28.4.334
AbstractAbstract PDFPubReaderePub

This study was performed to evaluate the actual temperature rise on the surface of Buchanan plugger using thermocouple. The heat carrier system 'System B Heatsource' (Model 1005, Analytic Technologies, Redmond, WA, USA) and the Buchanan pluggers of F, FM, M and ML sizes are used for this study. The temperature was set to 200℃ on digital display and the power level on it was set to 10. Five thermocouples were placed in direct contact with the surface of each size of Buchanan's pluggers at 1 mm increments from the tip to the 4 mm length of shank. The heat control spring was touched for 5 seconds, and the temperature rise on the surface of the pluggers were measured at 1 sec intervals for more than 5 seconds with an accuracy of 0.01 using Data Logger. The data were statistically analyzed by one-way ANOVA.

The results were as follows.

The position at which the temperature peaked was approximately at 1~2 mm far from the tip of Buchanan plugger (p<0.01).

The peak temperature was 215.25±2.28℃ in F plugger, 185.94±2.19℃ in FM plugger, 169.51±9.12℃ in M plugger, and 160.79±1.27℃ in ML plugger and the peak temperature was highest in F plugger and followed by, in descending order, FM plugger, M plugger. ML plugger showed the lowest peak temperature (p<0.01).

The temperature on the pluggers was decreased with the increase of touching time.

This results suggest that the actual temperature on the surface of the pluggers does not correlate well with the temperature set on digital display. Heat concentrates around the tip. The larger plugger reveals lower temperature rise relatively.

  • 15 View
  • 0 Download
Close layer
Analysis of temperature rise on the root surface during continuous wave of condensation technique
Young-Ju Kim, Yun-Chan Hwang, Sun-Ho Kim, In-Nam Hwang, Bo-Young Choi, Young-Jin Jeong, Woo-Nam Juhng, Won-Mann Oh
J Korean Acad Conserv Dent 2003;28(4):341-347.   Published online July 31, 2003
DOI: https://doi.org/10.5395/JKACD.2003.28.4.341
AbstractAbstract PDFPubReaderePub

This study was conducted to evaluate the temperature rise on the root surface while the root canal is being obturated using continuous wave of condensation technique. Maxillary central incisor was prepared for repeated canal obturation. Ten thermocouples (Omega Engineering Inc., Stanford, USA) were placed at 1 mm increment from the anatomical root apex. The real temperature of Buchanan plugger was recorded before insertion into the root canal. The root canal was obturated with continuous wave of condensation technique as described by Buchanan and the root surface temperature was recorded during obturation at 150℃, 200℃, 250℃ and 300℃ temperature settings of System B HeatSource (Model 1005, Analytic technologies, Redmond, WA, USA). After completion of the temperature recording, the dentinal-cementum thickness at each sites was measured. The data were analyzed using one-way ANOVA followed by Scheffe' s test and linear regression test.

The results were as follows.

When the temperature was set at 150℃, 200℃, 250℃ and 300℃ on the digital display of System B HeatSource, the real temperature of the plugger at the 1mm point from the tip revealed 130.82±2.96℃, 158.00±5.26℃, 215.92±6.91℃ and 249.88±3.65℃ respectively.

The position of 8 mm from the anatomical apex showed the highest temperature increase at each temperature settings and it was significantly higher than those of other positions (p<0.01). The temperature rise was constantly increased toward coronal portion from apex of the root.

The maximum temperature increase on the root surface was 2.37±0.09℃ at 150℃ setting, 3.11±0.12℃ at 200℃ setting, 3.93±0.09℃ at 250℃ setting and 5.69±0.15℃ at 300℃ setting respectively.

These results suggest that it be relatively kind to the supporting tissues of the root that the root canal is obturated using continuous wave of condensation technique at 150℃, 200℃, 250℃ and 300℃ temperature settings on digital temperature display of System B HeatSource.

  • 20 View
  • 0 Download
Close layer
Infrared thermographic analysis of temperature rise on the surface of buchanan plugger
Sung-A Choi, Sun-Ho Kim, Yun-Chan Hwang, Chang Youn, Byung-Ju Oh, Bo-Young Choi, Woo-Nam Juhng, Sun-Wa Jeong, In-Nam Hwang, Won-Mann Oh
J Korean Acad Conserv Dent 2002;27(4):370-381.   Published online July 31, 2002
DOI: https://doi.org/10.5395/JKACD.2002.27.4.370
AbstractAbstract PDFPubReaderePub

This study was performed to evaluate the temperature rise on various position of the Buchanan plugger, the peak temperature of plugger's type and the temperature change by its touching time of heat control spring.

The heat carrier system 'System B'(Model 1005, Analytic Technologies, USA) and the Buchanan's pluggers of F, FM, M and ML sizes are used for this study. The temperature was set to 200℃ which Dr. Buchanan's "continuous wave of condensation" technique recommended on digital display and the power level on it was set to 10. In order to apply heat on the Buchanan's pluggers, the heat control spring was touched for 1, 2, 3, 4 and 5 seconds respectively. The temperature rise on the surface of the pluggers were measured at 0.5 mm intervals from tip to 20 mm length of shank using the infrared thermography (Radiation Thermometer-IR Temper, NEC San-ei Instruments, Ltd, Japan) and TH31-702 Data capture software program (NEC San-ei Instruments, Ltd, Japan). Data were analyzed using a one way ANOVA followed by Duncan's multiple range test and linear regression test.

The results as follows.

1. The position at which temperature peaked was approximately at 0.5 mm to 1.5 mm far from the tip of Buchanan's pluggers (p<0.001). The temperature was constantly decreased toward the shank from the tip of it (p<0.001).

2. When the pluggers were heated over 5 seconds, the peak temperature by time of measurement revealed from 253.3±10.5℃ to 192.1±3.3℃ in a touch for 1 sec, from 218.6±5.0℃ to 179.5±4.2℃ in a touch for 2 sec, from 197.5±3.0℃ to 167.6±3.7℃ in a touch for 3 sec, from 183.7±2.5℃ to 159.8±3.6℃ in a touch for 4 sec and from 164.9±2.0℃ to 158.4±1.8℃ in a touch for 5 sec. A touch for 1 sec showed the highest peak temperature, followed by, in descending order, 2 sec, 3 sec, 4 sec. A touch for 5 sec showed the lowest peak temperature (p<0.001).

3. A each type of pluggers showed different peak temperatures. The peak temperature was the highest in F type and followed by, in descending order, M type, ML type. FM type revealed the lowest peak temperature (p<0.001).

The results of this study indicated that pluggers are designed to concentrate heat at around its tip, its actual temperature does not correlate well with the temperature which Buchanan's "continuous wave of condensation" technique recommend, and finally a quick touch of heat control spring for 1sec reveals the highest temperature rise.