Skip Navigation
Skip to contents

Restor Dent Endod : Restorative Dentistry & Endodontics

OPEN ACCESS

Author index

Page Path
HOME > Browse articles > Author index
Search
Jin-Suk Cho 1 Article
Analysis of temperature rise on the surface of buchanan plugger using thermocouple
Jin-Suk Cho, Yun-Chan Hwang, Sun-Ho Kim, In-Nam Hwang, Bo-Young Choi, Young-Jin Jeong, Woo-Nam Juhng, Won-Mann Oh
J Korean Acad Conserv Dent 2003;28(4):334-340.   Published online July 31, 2003
DOI: https://doi.org/10.5395/JKACD.2003.28.4.334
AbstractAbstract PDFPubReaderePub

This study was performed to evaluate the actual temperature rise on the surface of Buchanan plugger using thermocouple. The heat carrier system 'System B Heatsource' (Model 1005, Analytic Technologies, Redmond, WA, USA) and the Buchanan pluggers of F, FM, M and ML sizes are used for this study. The temperature was set to 200℃ on digital display and the power level on it was set to 10. Five thermocouples were placed in direct contact with the surface of each size of Buchanan's pluggers at 1 mm increments from the tip to the 4 mm length of shank. The heat control spring was touched for 5 seconds, and the temperature rise on the surface of the pluggers were measured at 1 sec intervals for more than 5 seconds with an accuracy of 0.01 using Data Logger. The data were statistically analyzed by one-way ANOVA.

The results were as follows.

The position at which the temperature peaked was approximately at 1~2 mm far from the tip of Buchanan plugger (p<0.01).

The peak temperature was 215.25±2.28℃ in F plugger, 185.94±2.19℃ in FM plugger, 169.51±9.12℃ in M plugger, and 160.79±1.27℃ in ML plugger and the peak temperature was highest in F plugger and followed by, in descending order, FM plugger, M plugger. ML plugger showed the lowest peak temperature (p<0.01).

The temperature on the pluggers was decreased with the increase of touching time.

This results suggest that the actual temperature on the surface of the pluggers does not correlate well with the temperature set on digital display. Heat concentrates around the tip. The larger plugger reveals lower temperature rise relatively.

  • 15 View
  • 0 Download
Close layer

Restor Dent Endod : Restorative Dentistry & Endodontics
Close layer
TOP