Effect of internal stress on cyclic fatigue failure in K3
Skip Navigation
Skip to contents

Restor Dent Endod : Restorative Dentistry & Endodontics

OPEN ACCESS

Articles

Page Path
HOME > Restor Dent Endod > Volume 37(2); 2012 > Article
Research Article Effect of internal stress on cyclic fatigue failure in K3
Jun-Young Kim, DDS, MSD, Jin-Woo Kim, DDS, PhD, Kyung-Mo Cho, DDS, PhD, Se-Hee Park, DDS, PhD
Restorative Dentistry & Endodontics 2012;37(2):74-78.
DOI: https://doi.org/10.5395/rde.2012.37.2.74
Published online: May 18, 2012

Department of Conservative Dentistry, Gangneung-Wonju National University School of Dentistry, Gangneung, Korea.

Correspondence to Se-Hee Park, DDS, PhD. Assistant Professor, Department of Conservative Dentistry, Gangneung-Wonju National University School of Dentistry, 120 Daehang-ro, Gangneung, Korea 210-702. TEL, +82-33-640-2760; FAX, +82-33-640-3103; drendo@gwnu.ac.kr
• Received: December 22, 2011   • Revised: February 6, 2012   • Accepted: February 12, 2012

©Copyights 2012. The Korean Academy of Conservative Dentistry.

  • 17 Views
  • 0 Download
prev next
  • Objectives
    This study aimed to evaluate the relationship between the cyclic fatigue of a K3 file and internal stress intentionally induced until the activation of the auto-stop function of the torque-controlled motor.
  • Materials and Methods
    K3 (Sybron Endo) .04 and .06 taper, size 25, 30, 35, 40 and 45 were used in this study. To give the internal stress, the K3 files were put into the .02 taper Endo-Training-Bloc (Dentsply Maillefer) until the activation of the auto-stop function of the torque-controlled motor. The rotation speed was 300 rpm and torque value was 1.0 N·cm. K3 were grouped by the number of induced internal stress and randomly distributed to 4 experimental groups (n = 10, Stress 0 [control], Stress 1, Stress 2 and Stress 3). For measuring the cyclic fatigue failure, the K3 files were worked against a sloped glass block and time for file separation was recorded. Data was statistically analyzed Statistical analyses were performed using two-way ANOVA and Duncan post-hoc test at p < 0.05 level.
  • Results
    Except .04 taper size 30 in Stress 1 group, there were statistically significant differences in time for file separation between control and all experimental groups. K3 with .04 taper showed higher cyclic fatigue resistance than those of .06 taper.
  • Conclusion
    In the limitation of this study, the cyclic fatigue of the K3 file was influenced by the accumulated internal stress from use until the auto-stop function was activated by the torque-controlled motor. Therefore, clinicians should avoid the reuse of the K3 file that has undergone auto-stops.

This study was supported by research funding of 2011 year from Gangneung-Wonju National University Dental Hospital (2011-0089).

No potential conflict of interest relevant to this article was reported.

  • 1. Walia HM, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of Nitinol root canal files. J Endod. 1988;14: 346-351.ArticlePubMed
  • 2. Gambill JM, Alder M, del Rio CE. Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography. J Endod. 1996;22: 369-375.ArticlePubMed
  • 3. Glossen CR, Haller RH, Dove SB, del Rio CE. A comparison of root canal preparations using Ni-Ti hand, Ni-Ti engine-driven, and K-Flex endodontic instruments. J Endod. 1995;21: 146-151.PubMed
  • 4. Elliott LM, Curtis RV, Pitt Ford TR. Cutting pattern of nickel-titanium files using two preparation techniques. Endod Dent Traumatol. 1998;14: 10-15.ArticlePubMed
  • 5. Cheung GSP. Instrument fracture: mechanisms, removal of fragments, and clinical outcomes. Endodontic Topics. 2009;16: 1-26.Article
  • 6. Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT Series X rotary nickel-titanium files. J Endod. 2009;35: 404-407.ArticlePubMed
  • 7. Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod. 1997;23: 77-85.PubMed
  • 8. Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26: 161-165.ArticlePubMed
  • 9. Kuhn G, Tavernier B, Jordan L. Influence of structure on nickel-titanium endodontic instruments failure. J Endod. 2001;27: 516-520.ArticlePubMed
  • 10. Gambarini G. Advantages and disadvantages of new torque-controlled endodontic motors and low-torque NiTi rotary instrumentation. Aust Endod J. 2001;27: 99-104.ArticlePubMed
  • 11. Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod. 2004;30: 228-230.PubMed
  • 12. Barbosa FO, Gomes JA, de Araújo MC. Fractographic analysis of K3 nickel-titanium rotary instruments submitted to different modes of mechanical loading. J Endod. 2008;34: 994-998.ArticlePubMed
  • 13. Li UM, Lee BS, Shih CT, Lan WH, Lin CP. Cyclic fatigue of endodontic nickel titanium rotary instruments: static and dynamic tests. J Endod. 2002;28: 448-451.ArticlePubMed
  • 14. Park SH, Seo HW, Hong CU. An evaluation of rotational stability in endodontic electronic motors. J Korean Acad Conserv Dent. 2010;35: 246-256.Article
  • 15. Melo MC, Pereira ES, Viana AC, Fonseca AM, Buono VT, Bahia MG. Dimensional characterization and mechanical behaviour of K3 rotary instruments. Int Endod J. 2008;41: 329-338.PubMed
  • 16. Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod. 2009;35: 1469-1476.ArticlePubMed
  • 17. Bahia MG, Buono VT. Decrease in the fatigue resistance of nickel-titanium rotary instruments after clinical use in curved root canals. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;100: 249-255.ArticlePubMed
  • 18. Haikel Y, Gasser P, Allemann C. Dynamic fracture of hybrid endodontic hand instruments compared with traditional files. J Endod. 1991;17: 217-220.ArticlePubMed
  • 19. Haïkel Y, Serfaty R, Bateman G, Senger B, Allemann C. Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments. J Endod. 1999;25: 434-440.PubMed
  • 20. Cheung GS, Darvell BW. Fatigue testing of a NiTi rotary instrument. Part 1: strain-life relationship. Int Endod J. 2007;40: 612-618.ArticlePubMed
  • 21. Schäfer E, Dzepina A, Danesh G. Bending properties of rotary nickel-titanium instruments. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;96: 757-763.ArticlePubMed
  • 22. Ullmann CJ, Peters OA. Effect of cyclic fatigue on static fracture loads in ProTaper nickel-titanium rotary instruments. J Endod. 2005;31: 183-186.ArticlePubMed
  • 23. Shin YM, Kim ES, Kim KM, Kum KY. Effect of surface defects and cross-sectional configuration on the fatigue fracture of NiTi rotary files under cyclic loading. J Korean Acad Conserv Dent. 2004;29: 267-272.
  • 24. Lee JK, Kim ES, Kang MW, Kum KY. The effect of surface defects on the cyclic fatigue fracture of HEROShaper Ni-Ti rotary files in a dynamic model: a fractographic analysis. J Korean Acad Conserv Dent. 2007;32: 130-137.Article
Figure 1
Representative image of cyclic fatigue test (inclined glass plane).
rde-37-74-g001.jpg
Figure 2
FE-SEM view of K3. (a) Torsional fracture of K3 .04/#25. Central dimples and circular abrasion mark (white arrow) (×200); (b) Torsional distorsion of K3 .04/#25; (c) Torsional distorsion of K3 .04/#30 (×30); (d) Cyclic fatigue fracture of K3 .04/#45. Radial striations linked at a point on the edge (black arrow) (×120). The surfaces of cyclic fatigue fracture in all experimental groups showed same patterns. FE-SEM, field emission scanning electron microscope.
rde-37-74-g002.jpg
Table 1
Time for file separation of K3 .04 taper

Same superscripts shows no statistically significant difference in same size.

Stress 0 group was not induced internal stress, Stress 1 group was induced internal stress 1 time, Stress 2 was induced internal stress 2 times, and Stress 3 was induced internal stress 3times by activation of the auto-stop function of the torque-controlled motor.

rde-37-74-i001.jpg
Table 2
Time for file separation of K3 .06 taper

Same superscripts shows no statistically significant difference in same size.

Stress 0 group was not induced internal stress, Stress 1 group was induced internal stress 1 time, Stress 2 was induced internal stress 2 times, and Stress 3 was induced internal stress 3times by activation of the auto-stop function of the torque-controlled motor.

rde-37-74-i002.jpg

Tables & Figures

REFERENCES

    Citations

    Citations to this article as recorded by  

      • ePub LinkePub Link
      • Cite
        CITE
        export Copy Download
        Close
        Download Citation
        Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

        Format:
        • RIS — For EndNote, ProCite, RefWorks, and most other reference management software
        • BibTeX — For JabRef, BibDesk, and other BibTeX-specific software
        Include:
        • Citation for the content below
        Effect of internal stress on cyclic fatigue failure in K3
        Restor Dent Endod. 2012;37(2):74-78.   Published online May 18, 2012
        Close
      • XML DownloadXML Download
      Figure
      • 0
      • 1
      Effect of internal stress on cyclic fatigue failure in K3
      Image Image
      Figure 1 Representative image of cyclic fatigue test (inclined glass plane).
      Figure 2 FE-SEM view of K3. (a) Torsional fracture of K3 .04/#25. Central dimples and circular abrasion mark (white arrow) (×200); (b) Torsional distorsion of K3 .04/#25; (c) Torsional distorsion of K3 .04/#30 (×30); (d) Cyclic fatigue fracture of K3 .04/#45. Radial striations linked at a point on the edge (black arrow) (×120). The surfaces of cyclic fatigue fracture in all experimental groups showed same patterns. FE-SEM, field emission scanning electron microscope.
      Effect of internal stress on cyclic fatigue failure in K3

      Time for file separation of K3 .04 taper

      Same superscripts shows no statistically significant difference in same size.

      Stress 0 group was not induced internal stress, Stress 1 group was induced internal stress 1 time, Stress 2 was induced internal stress 2 times, and Stress 3 was induced internal stress 3times by activation of the auto-stop function of the torque-controlled motor.

      Time for file separation of K3 .06 taper

      Same superscripts shows no statistically significant difference in same size.

      Stress 0 group was not induced internal stress, Stress 1 group was induced internal stress 1 time, Stress 2 was induced internal stress 2 times, and Stress 3 was induced internal stress 3times by activation of the auto-stop function of the torque-controlled motor.

      Table 1 Time for file separation of K3 .04 taper

      Same superscripts shows no statistically significant difference in same size.

      Stress 0 group was not induced internal stress, Stress 1 group was induced internal stress 1 time, Stress 2 was induced internal stress 2 times, and Stress 3 was induced internal stress 3times by activation of the auto-stop function of the torque-controlled motor.

      Table 2 Time for file separation of K3 .06 taper

      Same superscripts shows no statistically significant difference in same size.

      Stress 0 group was not induced internal stress, Stress 1 group was induced internal stress 1 time, Stress 2 was induced internal stress 2 times, and Stress 3 was induced internal stress 3times by activation of the auto-stop function of the torque-controlled motor.