Skip Navigation
Skip to contents

Restor Dent Endod : Restorative Dentistry & Endodontics

OPEN ACCESS

Search

Page Path
HOME > Search
1 "C-Pilot"
Filter
Filter
Article category
Keywords
Publication year
Authors
Research Article
Comparison of the ability of Reciproc and Reciproc Blue instruments to reach the full working length with or without glide path preparation
Mehmet Adıguzel, Pelin Tufenkci
Restor Dent Endod 2018;43(4):e41.   Published online November 1, 2018
DOI: https://doi.org/10.5395/rde.2018.43.e41
AbstractAbstract PDFPubReaderePub
Objectives

The purpose of the present study was to compare the mean preparation times and frequency with which Reciproc and Reciproc Blue instruments reached the full working length in mandibular molars, with or without glide path preparation.

Materials and Methods

Previously untreated mesiobuccal and mesiolingual canals with completely formed apices were randomly divided into 6 groups (n = 50) depending on the usage of Reciproc (RC; VDW), Reciproc Blue (RC Blue; VDW), C-Pilot (CP; VDW), and R-Pilot (RP; VDW) files: RC, RC Blue, RC + C-Pilot (RC-CP), RC-Blue + C-Pilot (RC Blue-CP), RC+R-Pilot (RC-RP), and RC Blue + R-Pilot (RC Blue-RP). A glide path was prepared using the hand-operated C-Pilot or the machine-operated R-Pilot instruments, respectively. The χ2 test, analysis of variance, and the Tukey post hoc test were used for statistical comparisons.

Results

No statistically significant differences were observed in the distribution of the frequency of reaching the full working length in the RC (94%), RC Blue (88%), RC-CP (94%), RC Blue-CP (90%), RC-RP (96%), and RC Blue-RP (92%) groups (p > 0.05).

Conclusions

Preparation of a glide path did not have a significant effect on reaching the full working length using these systems.

  • 23 View
  • 0 Download
Close layer

Restor Dent Endod : Restorative Dentistry & Endodontics
Close layer
TOP