Analysis of temperature rise on the surface of buchanan plugger using thermocouple

Analysis of temperature rise on the surface of buchanan plugger using thermocouple

Article information

Restor Dent Endod. 2003;28(4):334-340
Publication date (electronic) : 2003 July 31
doi : https://doi.org/10.5395/JKACD.2003.28.4.334
Department of Conservative Dentistry, College of Dentistry, DSRI, Chonnam National University, Korea.
#School of Mechanical Systems Engineering, Chonnam National University, Korea.
Corresponding author (wmoh@chonnam.ac.kr)

Abstract

This study was performed to evaluate the actual temperature rise on the surface of Buchanan plugger using thermocouple. The heat carrier system 'System B Heatsource' (Model 1005, Analytic Technologies, Redmond, WA, USA) and the Buchanan pluggers of F, FM, M and ML sizes are used for this study. The temperature was set to 200℃ on digital display and the power level on it was set to 10. Five thermocouples were placed in direct contact with the surface of each size of Buchanan's pluggers at 1 mm increments from the tip to the 4 mm length of shank. The heat control spring was touched for 5 seconds, and the temperature rise on the surface of the pluggers were measured at 1 sec intervals for more than 5 seconds with an accuracy of 0.01 using Data Logger. The data were statistically analyzed by one-way ANOVA.

The results were as follows.

  1. The position at which the temperature peaked was approximately at 1~2 mm far from the tip of Buchanan plugger (p<0.01).

  2. The peak temperature was 215.25±2.28℃ in F plugger, 185.94±2.19℃ in FM plugger, 169.51±9.12℃ in M plugger, and 160.79±1.27℃ in ML plugger and the peak temperature was highest in F plugger and followed by, in descending order, FM plugger, M plugger. ML plugger showed the lowest peak temperature (p<0.01).

  3. The temperature on the pluggers was decreased with the increase of touching time.

This results suggest that the actual temperature on the surface of the pluggers does not correlate well with the temperature set on digital display. Heat concentrates around the tip. The larger plugger reveals lower temperature rise relatively.

References

1. Schilder H. Filling root canals in three dimensions. Dent Clin North Am 1967. 723–744.
2. Nguyen NT. Obturation of the Root Canal System. Pathways of the Pulp 1994. 6th edth ed. St. Louis: Mosby Co.; 219–271.
3. Jurcak JJ, Weller RN, Kulild JC, Donley DL. In vitro intracanal temperatures produced during warm lateral condensation of gutta-percha. J Endod 1992. 181–3.
4. Lee FS, Van Cura JE, BeGole E. A comparison of root surface temperatures using different obturation heat source. J Endod 1998. 24617–620.
5. Hand RE, Hugel EF, Tsaknis PJ. Effect of a warm gutta-percha technique on the lateral periodontium. Oral Surg Oral Med Oral Pathol 1976. 42395–401.
6. Marlin J, Schilder H. Physical properties of gutta-percha when subjected to heat and vertical condensation. Oral Surg Oral Med Oral Pathol 1973. 36872–879.
7. Barkhordar RA, Goodis HE, Watanabe L, Koundjian J. Evaluation of temperature rise on the outer surface of teeth during root canal obturation technique. Quintessence Int 1990. 21585–588.
8. Eriksson A, Albrektsson T, Grane B, McQueen D. Thermal injury to bone: a vital-microscopic description of heat effect. Int J Oral Surg 1982. 11115–121.
9. Eriksson RA, Albrektsson T. The effect of heat on bone regeneration. J Oral Maxillofac Surg 1984. 42705–711.
10. Eriksson AR, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 1983. 50101–107.
11. Romero AD, Green DB, Wucherpfennig AL. Heat transfer to the periodontal ligament during root obturation procedures using an in vitro model. J Endod 2000. 2685–87.
12. Weller RN, Jurcak JJ, Donley DL, Kulild JC. A new model system for measuring intracanal temperatures. J Endod 1991. 17491–494.
13. Floren JW, Weller RN, Pashley DH, Kimbrough WF. Changes in root surface temperatures with in vitro use of the System B HeatSource. J Endod 1999. 25593–595.
14. Choi SA, Kim SH, Hwang YC, Youn C, Oh BJ, Choi BY, Juhng WN, Jeong SW, Hwang IN, Oh WM. Infrared thermographic analysis of temperature rise on the surface of buchanan plugger. J Korean Acad Conserv Dent 2002. 27382–388.
15. Buchanan LS. Continuous wave of condensation technique. Dent Today 1996. 1560–62. 64–67.
16. Analytic Technology Corp. Instruction guidelines for system B HeatSource model 1005 1997. Glendora, California, USA:

Article information Continued

Fig. 1

Schematic illustration of the whole system.

Fig. 2

Schematic illustration of Buchanan plugger with thermocouple.

Fig. 3

The fixation device of Buchanan plugger with System B HeatSource and data logger.

Fig. 4

Temperature changes on the surface of Buchanan plugger at 1 sec of touching time (℃).

(setting temperature: 200℃, power level: 10).

*: significant difference among pluggers (p<0.01).

@: significant difference among distances (p<0.01).

Fig. 5

Temperature changes on the surface of Buchanan plugger at 2 sec of touching time (℃).

(setting temperature: 200℃, power level: 10)

*: significant difference among pluggers (p<0.01).

@: significant difference among distances (p<0.01).

Fig. 6

Temperature changes on the surface of Buchanan plugger at 3 sec of touching time (℃).

(setting temperature: 200℃, power level: 10)

*: significant difference among pluggers (p<0.01).

@: significant difference among distances (p<0.01).

Fig. 7

Temperature changes on the surface of Buchanan plugger at 4 sec of touching time (℃).

(setting temperature: 200℃, power level: 10)

*: significant difference among pluggers (p<0.01).

@: significant difference among distances (p<0.01).

Fig. 8

Temperature changes on the surface of Buchanan plugger at 5 sec of touching time(℃).

(setting temperature: 200℃, power level: 10)

*: significant difference among pluggers (p<0.01).

@: significant difference among distances (p<0.01).

Table 1

Temperature changes on the surface of Buchanan plugger (F) (Mean±S.D. (℃), n=10)

Table 1

*: significant difference among time of measurement (p<0.01).

@: significant difference among distances (p<0.01).

Table 2

Temperature changes on the surface of Buchanan plugger (FM) (Mean±S.D. (℃), n=10)

Table 2

*: significant difference among time of measurement (p<0.01).

@: significant difference among distances (p<0.01).

Table 3

Temperature changes on the surface of Buchanan plugger (M) (Mean±S.D. (℃), n=10)

Table 3

*: significant difference among time of measurement (p<0.01).

@: significant difference among distances (p<0.01).

Table 4

Temperature changes on the surface of Buchanan plugger (ML) (Mean±S.D. (℃), n=10)

Table 4

*: significant difference among time of measurement (p<0.01).

@: significant difference among distances (p<0.01).