Power density of light curing units through resin inlays fabricated with direct and indirect composites

Power density of light curing units through resin inlays fabricated with direct and indirect composites

Article information

Restor Dent Endod. 2010;35(5):353-358
Publication date (electronic) : 2010 September 30
doi : https://doi.org/10.5395/JKACD.2010.35.5.353
1Department of Conservative Dentistry, Wonkwang University College of Dentistry and Dental Research Institute, Iksan, Korea.
2Dental Laboratory, Wonkwang University Dental Hospital, Iksan, Korea.
Correspondence to Hoon-Sang Chang, DDS, PhD. Assistant Professor, Department of Conservative Dentistry, Wonkwang University College of Dentistry and Dental Research Institute, 344-2 Shinyong-dong, Iksan, Korea 570-210. Tel, +82-63-859-2931; Fax, +82-63-859-2932; husch03@wonkwang.ac.kr
Received 2010 July 22; Revised 2010 August 09; Accepted 2010 August 10.

Abstract

Objectives

The purpose of this study was to measure the power density of light curing units transmitted through resin inlays fabricated with direct composite (Filtek Z350, Filtek Supreme XT) and indirect composite (Sinfony).

Materials and Methods

A3 shade of Z350, A3B and A3E shades of Supreme XT, and A3, E3, and T1 shades of Sinfony were used to fabricate the resin inlays in 1.5 mm thickness. The power density of a halogen light curing unit (Optilux 360) and an LED light curing unit (Elipar S10) through the fabricated resin inlays was measured with a hand held dental radiometer (Cure Rite). To investigate the effect of each composite layer consisting the resin inlays on light transmission, resin specimens of each shade were fabricated in 0.5 mm thickness and power density was measured through the resin specimens.

Results

The power density through the resin inlays was lowest with the Z350 A3, followed by Supreme XT A3B and A3E. The power density was highest with Sinfony A3, E3, and T1 (p < 0.05). The power density through 0.5 mm thick resin specimens was lowest with dentin shades, Sinfony A3, Z350 A3, Supreme XT A3B, followed by enamel shades, Supreme XT A3E and Sinfony E3. The power density was highest with translucent shade, Sinfony T1 (p < 0.05).

Conclusions

Using indirect lab composites with dentin, enamel, and translucent shades rather than direct composites with one or two shades could be advantageous in transmitting curing lights through resin inlays.

References

1. Kramer N, Lohbauer U, Frankenberger R. Adhesive luting of indirect restorations. Am J Dent 2000. 13(Spec No)60D–76D.
2. McLean JW. The science and art of dental ceramics. Oper Dent 1991. 16149–156.
3. Touati B, Aidan N. Second generation laboratory composite resins for indirect restorations. J Esthet Dent 1997. 9108–118.
4. Cesar PF, Miranda WG Jr, Braga RR. Influence of shade and storage time on the flexural strength, flexural modulus, and hardness of composites used for indirect restorations. J Prosthet Dent 2001. 86289–296.
5. Soares CJ, Pizi EC, Fonseca RB, Martins LR. Mechanical properties of light-cured composites polymerized with several additional post-curing methods. Oper Dent 2005. 30389–394.
6. Borba M, Della Bona A, Cecchetti D. Flexural strength and hardness of direct and indirect composites. Braz Oral Res 2009. 235–10.
7. el-Mowafy O. The use of resin cements in restorative dentistry to overcome retention problems. J Can Dent Assoc 2001. 6797–102.
8. Sorensen JA, Kang SK, Avera SP. Porcelain-composite interface microleakage with various porcelain surface treatments. Dent Mater 1991. 7118–123.
9. Jensen ME, Sheth JJ, Tolliver D. Etched-porcelain resin-bonded full-veneer crowns:in vitro fracture resistance. Compendium 1989. 10336–338. 40–41. 44–47.
10. el-Badrawy WA, el-Mowafy OM. Chemical versus dual curing of resin inlay cements. J Prosthet Dent 1995. 73515–524.
11. Hasegawa EA, Boyer DB, Chan DC. Hardening of dual-cured cements under composite resin inlays. J Prosthet Dent 1991. 66187–192.
12. Rueggeberg FA, Jordan DM. Effect of light-tip distance on polymerization of resin composite. Int J Prosthodont 1993. 6364–370.
13. Prati C, Chersoni S, Montebugnoli L, Montanari G. Effect of air, dentin and resin-based composite thickness on light intensity reduction. Am J Dent 1999. 12231–234.
14. Fan PL, Schumacher RM, Azzolin K, Geary R, Eichmiller FC. Curing-light intensity and depth of cure of resin-based composites tested according to international standards. J Am Dent Assoc 2002. 133429–434.
15. Na JS, Jeong SW, Hwang YC, Kim SH, Yun C, Oh WM, et al. Effect of light source and shade on depth of cure of composites. J Korean Acad Conserv Dent 2002. 27561–568.
16. Park JJ, Park JW, Park SH, Park JM, Kwon TK, Kim SK. The effect of irradiation modes on polymerization and microleakage of composite resin. J Korean Acad Conserv Dent 2002. 27158–174.
17. Shin HJ, Kim JW, Cho KM. Polymerization ability of several light curing sources on composite resin. J Korean Acad Conserv Dent 2003. 28156–161.
18. McCabe JF, Carrick TE. Output from visible-light activation units and depth of cure of light-activated composites. J Dent Res 1989. 681534–1539.
19. Shortall AC. How light source and product shade influence cure depth for a contemporary composite. J Oral Rehabil 2005. 32906–911.
20. Campbell PM, Johnston WM, O'Brien WJ. Light scattering and gloss of an experimental quartz-filled composite. J Dent Res 1986. 65892–894.
21. Ruyter IE, Oysaed H. Conversion in different depths of ultraviolet and visible light activated composite materials. Acta Odontol Scand 1982. 40179–192.
22. dos Santos GB, Alto RV, Filho HR, da Silva EM, Fellows CE. Light transmission on dental resin composites. Dent Mater 2008. 24571–576.

Article information Continued

Table 1

Composition of the resin composites used in this study

Table 1

The composition of the organic matrix, filler, and filler contents was provided by the manufacturers.

Bis-GMA, Bisphenol A diglycidyl ether dimethacrylate; UDMA, urethane dimethacrylate; Bis-EMA, Bisphenol A polyethylene glycol diether dimethacrylate; TEGDMA, triethylene glycol dimethacrylate.

Table 2

Power density (mW/cm2 ± SD) of light curing units through resin specimens of 1.5 mm thickness

Table 2

*The superscripts with the same letters are not significantly different at α= 0.05.

Table 3

Power density (mW/cm2 ± SD) of light curing units through resin specimens of 0.5 mm thickness

Table 3

*The superscripts with the same letters are not significantly different at α= 0.05.